As-Needed/As-Required Maintenance
Replacing the CO
2
Membrane
UniCel DxC Systems Instructions For Use A13914
Maintenance
October 2005
Page 10-83
10
3
Center the membrane retainer (1) over the membrane clamp (3). Press the
membrane retainer into the clamp, thus securing the membrane (2).
Figure 10.58
4
Place the retainer assembly upside down on the work surface so the clamp
is on top and retainer on the bottom. Grasp the electrode carefully and
press it firmly against the membrane and into the membrane clamp.
Figure 10.59
Step
Action
, continued