UniCel DxC Systems Instr
u
ctions F
o
r Use A
13914
T
roub
leshooting Calibr
ation and Result Errors
October 2005
13-65
Common Error Messages and Corrective Actions
0x139F0069
9
Full CTS Tracking
is now available.
When you move a
tube to a different
instrument, it is not
necessary to
remove its cap.
STM: TS-CTS Tracking
System reconnect
detected.
Description: All of the systems that
we expect to be networked together
are connected (it has recovered
from a disconnect state).
• No operator action is necessary.
Databases will automatically
synchronize and caps can be left in
place.
0x13A00004
10
Requested Rack
Not Loaded
Rack: rack ID
WLM: Rack rack ID
cannot be loaded
(FileName:LineNumber)
Description: The sample rack
subsystem was unable to load a
rack, usually due to a motion error.
The rack will be removed and
processing will continue.
• Correct the accompanying motion
error and reload the rack.
0x13A0001C
10
Non-programmed
samples on rack.
WLM: Unprogrammed
samps,
rack: rack ID
[p:PositionSample ID]
[p:PositionSample ID]
[p:PositionSample ID]
[p:PositionSample ID]
Description: Rack has been loaded
with a sample which has not been
programmed.
• Make sure the complete sample bar
code is readable through the window in
the rack.
• Check communication with LIS.
• Manually program if necessary.
0x13A00022
10
The QC Database is
xx% full. Archiving
is suggested.
WLM: QC DB is xx%
percent full
Description: QC data base is
becoming full. When the database
becomes full, the oldest results will
be overwritten. To avoid this,
archive your QC data.
• Archive QC data so it will not be
overwritten.
Table 13.10
Error Messages and Corrective Actions
, continued
ID
Class #
Pop-up
Event Log
Description
Corrective Action