Common Error Messages and Corrective Actions
T
roub
leshooting Calibr
a
tion and Result Errors
UniCe
l DxC Systems Instr
u
ctions F
o
r Use A13914
13-46
October 2005
0x02680003
4
Sample tube did not
vent correctly.
Remove cap and
reload tube.
Sample ID: sample
id
Rack: rack id
Position: rack
position CC
Sample Probe
CAU: CC Sample Tube
Excess Vacuum Error.
Message ID:
0x02680003
Subsystem: RASched
Sample ID: sample id
Rack ID: rack id
Rack Position: rack
position
Description: Probe detected
excessive vacuum in thick stopper
closed tube on second attempt.
Sample is aborted.
• Uncap tube and rerun.
• If occurs consistently for many
samples, check cap piercer for broken
blade.
• Perform CC ODC Alignment.
0x02680004
4
Sample tube did not
vent correctly.
Remove cap and
reload tube.
Sample ID: sample
id
Rack: rack id
Position: rack
position CC
Sample Probe
CAU: CC Sample Tube
Excess Pressure Error.
Message ID:
0x02680004
Subsystem: RASched
Sample ID: sample id
Rack ID: rack id
Rack Position: rack
position
Description: Probe detected
excessive pressure in thick stopper
closed tube on second attempt.
Sample is aborted.
• Uncap tube and rerun.
• If occurs consistently for many
samples, check cap piercer for broken
blade.
• Perform CC ODC Alignment.
Table 13.10
Error Messages and Corrective Actions
, continued
ID
Class #
Pop-up
Event Log
Description
Corrective Action