Common Error Messages and Corrective Actions
T
roub
leshooting Calibr
a
tion and Result Errors
UniCe
l DxC Systems Instr
u
ctions F
o
r Use A13914
13-58
October 2005
0x05DC01FB
4
CC Reagent
Carousel Motion
Error
CAU: Homing Error for
CC Reagent Carousel.
Message ID:
0x05DC01FB,
Subsystem: boot,
homing command
Description: This error occurs
when there is any homing issue
with the Reagent Carousel
Subsystem. This includes the
following:
• Upper/Lower reagent decks/
motors.
• Both decks flag sensors.
• The magnetic door sensor.
• Have any reagent cartridges been
loaded recently? If so, check to make
sure cartridges are fully seated and
refrigerator door is latched.
• Press the STOP button and HOME the
system.
• Check Event Log for specific
RGTWHL subsystem errors.
• Contact Beckman Coulter Technical
Support Center for additional
information.
0x05DC01FD
2
CC Sample Mixer
Error
CAU: Homing Error: $s
Message ID:
0x05DC01FD
Description: This motion error is
contained within the Sample Mixer
Crane and will only occur during a
home. This can be either a vertical
or rotational error.
• Make sure mixer has the necessary
room to move
• Make sure that the wiring and cabling
are secured properly
• Make sure mixer paddle is seated
firmly in it's housing
• If mixer appears out of position,
perform rotary or vertical alignment
• HOME and restart the system.
• If motion error continues, contact
Beckman Coulter Technical Support
Center for additional information.
Table 13.10
Error Messages and Corrective Actions
, continued
ID
Class #
Pop-up
Event Log
Description
Corrective Action