Three-Month Maintenance
Replace Syringe Plungers
Maintenance
UniCel DxC Systems Instructions For Use A13914
Page 10-46
October 2005
2
The syringe rods (1) are now positioned for placement.
Figure 10.28
3
• Unscrew the barrel of the reagent syringe (left back wall) or barrels of
two sample syringes (back wall above modular cups) to release each
syringe.
• Unscrew the round plunger end (1) from the plunger rod adaptor (2) and
remove the syringe.
Figure 10.29
Step
Action
, continued
(2 of 3)
A015944P.EPS
1
A015945P.EPS
2
1