As-Needed/As-Required Maintenance
Replacing the Sodium Measure/Reference Electrode
UniCel DxC Systems Instructions For Use A13914
Maintenance
October 2005
Page 10-77
10
11
Examine the electrode tip. If an air bubble is seen inside the tip, hold the
electrode cable near the end and spin the electrode for a few seconds. The
bubble should be pushed back inside the electrode and remain there away
from the tip.
12
Using a lintless tissue, carefully and thoroughly dry the electrode sides and
electrode port.
NOTICE
Do not touch the electrode tip.
13
Install a new quad-ring (1) on the electrode.
Insert the sodium electrode into the flow cell. Insert retainer nut (2) and
turn clockwise until finger-tight.
Figure 10.52
14
To test for proper seating of electrode, gently pull on electrode body. The
electrode assembly should not move. If the electrode moves, remove it and
try to install it once again. If installation is difficult, check for an extra or
missing quad-ring.
15
Reposition the flow cell on mounting panel and tighten the retaining
screws.
16
Reconnect the electrode cable to the appropriate connector.
17
Unclamp Line #22.
Step
Action
, continued
(4 of 4)