Reserved Racks/Obstruct Detect
Introduction
UniCel DxC Systems Instructions For Use A13914
System Setup Options
October 2005
Page 4-35
4
Reserved Racks/Obstruct Detect
Introduction
The Reserved Rack/Obstruction Detection option allows the operator to:
• Enable/disable the CTS (Cap Piercer) option.
• Enable/disable obstruction detection
• Assign reserved rack numbers for small sample volume containers
• Reserve a rack number or numbers for the analysis of HbA1c or IBCT samples.
• For a CTS system, assign rack numbers to the proper rack size.
• Restore defaults.
Setup
From the Setup screen, scroll down and select <17> Reserved Racks/Obstruction
Detection to enable/disable CTS, enable/disable obstruction detection, assign reserved
rack numbers for small sample volume containers, HbA1c or IBCT, or to restore
defaults.
The system must be in Standby or Stopped to make modifications. When changing the
CTS option, the operator must Stop and Home the system to completely enable the
feature.
The defaults are:
CTS:
Off
Obstruction Detection:
On
Reserved Racks:
None Reserved
Reserved Racks for
HbA1c or IBCT:
None Reserved