Monthly Maintenance
Replace CTS Blade (Sarstedt S-Monovette Type CTS Option only)
UniCel DxC Systems Instructions For Use A13914
Maintenance
October 2005
Page 10-35
10
4
Refer to
• Fully loosen the one captive blade screw (2).
• Fully loosen the two captive blade holder screws (1).
• Remove the blade holder.
Figure 10.19
5
• Hold the blade at the top of its shaft. Do NOT touch the points of the
blade.
• Lift up the blade to remove it.
• Discard the old blade into a biohazard, sharps container.
6
Note: The points of the blade are very sharp.
Remove a new blade from its container.
7
Hold the blade holder with the blade hole on the top. Note that the blade
hole is keyed and the blade has a slot.
Install the blade into the blade holder as follows:
• Rotate the blade, if necessary, until it falls into place.
• Finger tighten the captive blade screw.
• Remove the protection from the points of the blade.
• Do NOT touch the points of the blade.
Step
Action
, continued
(3 of 4)
A014850P.EPS
1
2