Host Communications
Introduction
UniCel DxC Systems Instructions For Use A13914
System Setup Options
October 2005
Page 4-37
4
Host Communications
Introduction
following lists the Host communications parameters, the options of each,
and the default values.
Setup
From the Setup screen, scroll down and select <20> Host Communications to select
parameters or restore defaults.
The system must be in Standby or Stopped to make modifications.
Refer to UniCel DxC Synchron Clinical Systems Host Interface Specifications for
comprehensive documentation for the host parameters.
Table 4.5
Host Communications Parameter Table
Parameter
Options
Default
Device ID
0–99
0
Flow Control
Software (XON/XOFF)
Hardware (RTS/CTS)
No Flow
Software (XON/XOFF)
Data Transmission Mode Transmission OFF,
Unidirectional, Bidirectional,
Bidirectional with Query
Unidirectional
Interface
CX7 Compatible
LX20/DxC
CX7 Compatible
Data Bits
8, 7
8
Stop Bits
1, 2
1
Parity
None, Odd, Even
None
Baud Rate
300, 1200, 2400, 4800, 9600,
19200
9600
3 Digit Rack Number
Enable or Disable
Disable
Requery
Automatic Requery of Host
Disable
Host does NOT accept
either > or < symbols
Enable or Disable
Enable
NOTICE
For Flow Control, do not select the Hardware (RTS/CTS) option. With the
Hardware option, ESD interference may affect the performance.