Two-Month Maintenance
Replace CTS Blade/Wick (CTS Option only)
UniCel DxC Systems Instructions For Use A13914
Maintenance
October 2005
Page 10-37
10
Two-Month Maintenance
Replace CTS Blade/Wick (CTS Option only)
Refer to
. Replace the Cap Piercer blade and its CTS Auto-Gloss wick
approximately every two months. If the system runs >500 samples per day, replace the
blade and wick more frequently. The blade count records the number of caps pierced
by the blade. Refer to
CHAPTER 11, System Status and Commands
Figure 10.20 Cap Piercer Blade with CTS Auto-Gloss Wick
Material Required:
• Medium Phillips screwdriver
• Small Phillips screwdriver
• Medium straight-edge screwdriver
Replacement Procedure
In the following procedure, the old blade and wick is removed in Steps 1-5 and the
new blade and wick is installed in Steps 6-10.
CAUTION
The Cap Piercer blade is very sharp and has been exposed to potentially
biohazardous fluids. To prevent injury or exposure, do not touch the points
of the blade and wear gloves.
1. Blade Clamp
2. Blade
3. Points
4. Wick Clip
5. Wick (in clip housing)
A007367P.EPS