As-Needed/As-Required Maintenance
EIC Port Cleaning
Maintenance
UniCel DxC Systems Instructions For Use A13914
Page 10-90
October 2005
12
Using the syringe filled with 10% bleach solution, connect the syringe tip
over the flow cell exit port #23 (1).
Figure 10.67
13
Flush the 10% bleach solution through the flow cell exit port. This flushes
the area between the valve port and flow cell exit port. Allow the bleach
solution to sit for 1 minute.
Note: The 1-minute waiting period applies only to the 10% bleach solution,
and not to the following deionized water or 70% isopropyl alcohol.
14
Repeat the flushing procedure through the flow cell exit port using
deionized water.
15
Repeat the flushing procedure through the flow cell exit port using 70%
isopropyl alcohol.
16
Repeat the flushing procedure through the flow cell exit port using
deionized water.
17
Thoroughly dry the EIC and attached solenoid valve using a dry lintless
tissue. Wipe the drip tray with deionized water and lintless tissue.
Thoroughly dry the drip tray.
Step
Action
, continued
(5 of 7)