Processing Parameters
Second Inject: Add Time
UniCel DxC Systems Instructions For Use A13914
User Defined Reagents
October 2005
Page 9-11
9
Second Inject: Add Time
Add Time programs the reagent dispense time into the cuvette. (Refer to
.)
• The Second Inject may be added at the same time as First Inject or a minimum of 8
seconds after sample has been pipetted. (Time is entered in increments of 1 second;
however, the reagent will be added at the 8-second interval closest to the time
defined. Sample addition occurs at 1 second.)
• A positive number indicates a Second Inject reagent component is added after
sample add.
• An add time value of -180 will dispense reagent at the same time as First Inject
reagent component.
Table 9.18
Allowable Entries for Add Time for the Second Inject
-180, 9 to 738 seconds
NOTICE
When defining a UDR, the second inject will be automatically moved to the
third inject if only the first and second injects are defined. The inject time must
not equal the default value (-180).