As-Needed/As-Required Maintenance
Cleaning the ISE Drain
Maintenance
UniCel DxC Systems Instructions For Use A13914
Page 10-96
October 2005
8
Reinstall the cleaned drain tube, or install a new drain tube (P/N 472698).
Insert the drain tube down through the O-ring until it sits flush on the
bottom of the ISE drain.
Note: If the fit is too snug to insert the drain tube, it may be necessary to
lubricate the O-ring on the inside of the lower drain. Without removing the
O-ring, apply a thin coat of Silicone compound to the O-ring area inside
the lower drain, prior to inserting the drain tube.
9
Clean the ISE drain top using a lintless tissue soaked with deionized water.
Thoroughly wipe dry the ISE drain.
10
NOTICE
Do not pinch, damage, or disconnect the adjacent connector wires while
reinstalling the drain top.
Reinstall the ISE drain top onto the drain. Align the two posts extending
from the bottom of the drain top, with the holes in the acrylic drain mount,
and gently lower the drain top until it sits flush (with no gaps) on the
mount.
Tighten the screw at drain top.
11
Select <OK> to exit the pop-up Maintenance Confirmation window when
procedure is complete.
12
After priming is complete, select <Close> in the pop-up Results window to
exit the ISE Service screen.
CAUTION
Pinching Hazard. When lowering the ISE module, keep fingers clear.
Lower the ISE module and close ISE module cover.
13
Calibrate all the ISE chemistries.
Step
Action
, continued
(3 of 3)