Common Error Messages and Corrective Actions
T
roub
leshooting Calibr
a
tion and Result Errors
UniCe
l DxC Systems Instr
u
ctions F
o
r Use A13914
13-48
October 2005
0x032B0005
2
CC Obstruction
Detection
Transducer Failure
CAU: CC Obstruction
Detection Transducer
Failure
Message ID:
0x032B0005
Subsystem: rasdSmc
Description: This error usually
occurs whenever the obstruction
transducer does not see fluid during
runtime. Often this comes from a
clogged degasser or lack of
Degassed water to the system. The
obstruction detection transducer is
looking for a certain range of
pressure during its primes before
sample aspiration. If it does not see
the correct differential, it will flag a
transducer error.
• Check Obstruction Detection
Transducer for obvious leaks or
dripping from connections or the
device itself and CC Sample Probe.
• Look for pinch in Water line (from
syringe to to top of probe).
• Make sure the CC sample Syringe is
properly seated.
• Make sure CC Sample Probe tubing is
properly connected at top of Probe.
• Contact Beckman Coulter Technical
Support Center for additional
information.
0x03350010
2
MC Obstruction
Detection
Transducer Failure.
CAU: MC Obstruction
Detection Transducer
Failure.
Message ID:
0x03350010
Subsystem: ppsdSmc
Description: MC Obstruction
detection hardware failure.
• Consistent failures may indicate issues
with the MC Obstruction Detection
Transducer.
• Contact Beckman Coulter Technical
Support Center for additional
information.
Table 13.10
Error Messages and Corrective Actions
, continued
ID
Class #
Pop-up
Event Log
Description
Corrective Action