leshooting Calibr
ation and Result Errors
October 2005
13-55
Common Error Messages and Corrective Actions
0x057A0009
4
Duplicate Rack ID
rack #
CAU: Duplicate Rack
ID
Message ID:
0x057A0009
Subsystem: SRShttl
Rack ID: rack #
Description: This error occurs
during a load when a rack ID is read
by the bar code reader and an
identical rack ID is already present
on the system. The system cannot
use duplicate ID racks
simultaneously.
• Compare the number of the rack that
was rejected with the racks currently
on the analyzer by looking at the Main
Screen.
• Check the actual bar code that is on the
rack correlates with the number that
the system is rejecting.
• Switch samples to a rack with a
different ID or wait until the samples in
the previous rack has completed testing
and has been offloaded.
• If the Main Screen shows the Rack, but
it is physically not on the instrument,
use the "Unload All" feature to clear
the sample wheel.
Table 13.10
Error Messages and Corrective Actions
, continued
ID
Class #
Pop-up
Event Log
Description
Corrective Action