As-Needed/As-Required Maintenance
Replacing the Chloride Electrode Tip
Maintenance
UniCel DxC Systems Instructions For Use A13914
Page 10-70
October 2005
Step
Action
1
Loosen the flow cell retaining screw (1) located at the top of the flow cell.
This is a captive screw and cannot be removed from the flow cell. Pull
assembly forward and upward to remove.
NOTICE
Do not disconnect any of the reagent lines attached to the flow cell.
Removal of the flow cell is not necessary for replacing electrodes
located on the right side of the flow cell.
Figure 10.43
2
Locate the Chloride electrode (1). Remove electrode by turning the
electrode retaining nut counterclockwise.
Figure 10.44
(1 of 3)
A015946P.EPS