CTS Wick Replacement Procedure
Clean the Blade
Maintenance
UniCel DxC Systems Instructions For Use A13914
Page 10-126
October 2005
Clean the Blade
It is recommended to clean the Blade at or near the instrument. If required, use the tray
to carry blade/wick items to other locations. Clean the Blade as follows:
Step
Action
1
Refer to
. Rinse the Blade with hot water as shown.
CAUTION
The points on the end of the Blade are very sharp and have been
exposed to potentially biohazardous fluids. To prevent possible injury
or exposure, avoid the Blade area.
Figure 10.85
(1 of 2)
A007560P.EPS