Requesting a Calibration
Calibrating DxC Chemistries Using Assigned Calibrator Racks
UniCel DxC Systems Instructions For Use A13914
Reagent Load/Calibration
October 2005
Page 5-19
5
Calibrating DxC Chemistries Using Assigned Calibrator Racks
From the Rgts/Cal screen, check the Calibration Time Left column shown in
days:hours:minutes and the Cal Status column to see which chemistries require
calibration.
Canceling a Calibration Request
Follow the steps below to cancel a calibration request.
Step
Action
1
Select the Rgts/Cal icon from the menu bar.
2
Select the chemistries to be calibrated. If necessary, use the up and down
arrows to select more chemistries.
3
Select Cal [F4].
4
Select List [F5] to review the calibrator load list. Make sure the lot number
on the calibrator bottle is the same as the lot number on the screen.
5
Get the assigned calibrator racks listed on the screen and place the cups
into assigned rack positions as shown on the load list.
6
Place calibrator into sample cups.
• For Aqua cals 1, 2, and 3, fill a 0.5 mL cup.
• All other calibrators need 3 drops of calibrator in a 0.5 mL cup.
7
Place the rack (1) in the autoloader (2) with the rack bar code label to the
right. Press the RUN button (3) on the analyzer. Refer to
.
Step
Action
1
Select the Rgts/Cal icon from the menu bar.
2
Select the appropriate requested and assigned chemistries to be canceled.
3
Select No Cal [F8] to cancel calibration request.
Chemistries currently selected and programmed for calibration may be
canceled only if the calibration is not yet in progress.
Calibration reports for a given calibrator do not print until all calibrations
are complete for each chemistry requested, unlike sample and control
results, which print on a cup-by-cup basis.