![Beckman Coulter UniCel DxC Synchron 600 Скачать руководство пользователя страница 480](http://html.mh-extra.com/html/beckman-coulter/unicel-dxc-synchron-600/unicel-dxc-synchron-600_instructions-for-use-manual_2719768480.webp)
Common Error Messages and Corrective Actions
T
roub
leshooting Calibr
a
tion and Result Errors
UniCe
l DxC Systems Instr
u
ctions F
o
r Use A13914
13-60
October 2005
0x05DC0202
2
MC Cups/ISE
Subsystem Error
CAU: Homing Error: $s
Message ID:
0x05DC01202
Description: This generic homing
error covers all aspects of the MC
side in the homing routine. This
error has several different triggers
ranging from motion issues to
"Attach" and download failures.
• If the event log parameter is
AttachPhos, AttachAlb,
AttachTP, AttachCrea,
AttachBun, AttachGlu or
AttachIse then the error occurred
attaching to one of the MC
modules.
• If the parameter is InitCupsIse
or PrimeCupsIse, the error
occurred during a cup or ISE
initialize or prime for a stop or
cold home.
• If the parameter is
WHomeMCIsePpsd or
WHomeMCCups then the error
occurred during a warm home of
the ISE, MC Sample Delivery
subsystem or MC cups.
System "Stopped". Generic homing
error message. Check Event Log for
specific subsystem errors.
• Press the STOP button and HOME the
system.
• Reboot system.
• Contact Beckman Coulter Technical
Support Center for additional
information.
Table 13.10
Error Messages and Corrective Actions
, continued
ID
Class #
Pop-up
Event Log
Description