Requesting a Calibration
Calibrating DxC Chemistries With Bar Code Labels
Reagent Load/Calibration
UniCel DxC Systems Instructions For Use A13914
Page 5-18
October 2005
Requesting a Calibration
Calibrating DxC Chemistries With Bar Code Labels
From the Rgts/Cal screen, check the Calibration Time Left column shown in
days:hours:minutes and the Cal Status column to see which chemistries require
calibration.
Step
Action
1
Select the Rgts/Cal icon from the menu bar.
2
Select the chemistries to be calibrated. If necessary, use the up and down
arrows to select more chemistries.
3
Select Cal [F4].
4
Select List [F5] to review the calibrator load list. Make sure the lot
number on the calibrator bottle agrees with the lot number seen on the
screen. Make sure the correct bar code labels are in the correct positions.
5
Place a calibrator bar code label (with the name of the calibrator to be
used) on a 16 mm transport tube. Repeat for all calibrators.
6
Place calibrator into sample cups.
• For Aqua cals 1, 2, and 3, fill a 0.5 mL cup.
7
Place the rack (1) in the autoloader (2) with the rack bar code label to the
right. Press the RUN button (3) on the analyzer.
Figure 5.9
A015925P.EPS
1
3