UniCel DxC Systems Instr
u
ctions F
o
r Use A
13914
T
roub
leshooting Calibr
ation and Result Errors
October 2005
13-33
Common Error Messages and Corrective Actions
0x01FA0016
4
Primary vacuum
level is low.
CAU: Primary vacuum
is low
Description: This error occurs
when the Primary vacuum drops
below 18in Hg for more than 15
seconds. There is no vacuum level
adjustment now and cannot be set.
• Check vacuum levels and HOME the
system.
• Make sure No-Foam is present, replace
if necessary.
• Make sure caps on pressurized bottles
are tight.
• Make sure Vacuum Accumulator
canister in the rear of the Hydro is
empty.
• Contact Beckman Coulter Technical
Support Center for additional
information.
0x01FA0019
4
10 PSI Air Supply
Pressure Is Low.
CAU: 10 PSI Air
Pressure Is Low.
Message ID:
0x01FA0019
Description: This error occurs
when the transducer line #129 is
outputting a value into the Hydro
Board that is below the 10 psi
specification.
• Check pressure levels and HOME the
system.
• Adjust pressure levels if necessary.
First, adjust pressure in Main system
above 17 psi. Then adjust the high
pressure system to 17 psi. Finally,
adjust the low pressure system. The
pressure range for the low pressure
system is 8–11 psi. Set it at 10 psi.
• Contact Beckman Coulter Technical
Support Center for additional
information.
Table 13.10
Error Messages and Corrective Actions
, continued
ID
Class #
Pop-up
Event Log
Description
Corrective Action