As-Needed/As-Required Maintenance
Flow Cell Flushing (Manual)
UniCel DxC Systems Instructions For Use A13914
Maintenance
October 2005
Page 10-97
10
Flow Cell Flushing (Manual)
Flushing the flow cell is recommended for troubleshooting drift, back-to-back errors
and range errors that may be caused by plugs or debris lodged in the electrode ports.
The following items should be available before starting this procedure:
• Flow cell flush kit (P/N 473181)
• 20 mL syringe
• hemostat
CAUTION
To prevent damage due to electrical static discharge (ESD), wear the wrist
ground strap when performing this procedure. For further instructions, refer
to "
" in the beginning of this chapter.
Step
Action
1
Remove ISE module cover and raise the ISE module to a locked position.
2
Prepare for ISE Maintenance and disable autoprime by selecting:
• the Utils icon from the menu bar,
• <2> Maintenance,
• <3> ISE Service.
Select <Continue> to drain the flow cell and prepare for maintenance.
3
Use a hemostat to clamp off line #35 at the top of the flow cell. Line the
area under the flow cell with absorbent tissue.
(1 of 3)