How to Use Reserved Racks
When NOT to Use a Reserved Rack
Preparing Samples for Analysis
UniCel DxC Systems Instructions For Use A13914
Page 3-8
October 2005
How to Use Reserved Racks
When NOT to Use a Reserved Rack
When you do CTS (Closed Tube Sampling) do NOT run a closed tube in a reserved
rack.
Recommended Use of Reserved Racks
Green Racks
Series 350–400 to run:
• nesting cups
• capillary collection containers
• Beckman Coulter Microtubes™
Gray Racks
Series 601 to run:
• "Barcoded" Calibrators and Controls
Purple Racks
• Series 401–420 to run HbA1c
• 0.5 mL Cup Insert (P/N 476399)
Brown Racks
• Series 451–499 to run IBCT
Blue Racks
0.5 mL Cup Insert (P/N 476399)
NOTICE
To identify a reserved rack, put a red ® on the top of the rack.