Reagent Load
Load/Unload Reagent
Reagent Load/Calibration
UniCel DxC Systems Instructions For Use A13914
Page 5-10
October 2005
4
To unload reagent:
When the screen prompts you to remove the cartridge (1), open the reagent
carousel door. The cartridge to unload will be positioned on either the
upper or lower reagent carousel. A blinking bar code reader indicates
which carousel position to unload from the upper and lower deck. Upper
deck positions (2) are 31 to 59. Lower deck positions (3) are 1 to 30.
Grasp the cartridge at top and bottom, lift up a little and pull back. Refer to
As the cartridge is moved in front of the bar code reader, a beep will
indicate that the bar code is read. Check the screen to verify reagent data is
removed from the position selected. If the bar code does not read,
manually clear the position. To manually clear, select the Chem to be
removed and select Clear [F1].
Figure 5.8
5
To load reagent:
When the screen prompts you to load the cartridge, open the reagent
carousel door. Move the cartridge in front of the bar code reader. A beep
will indicate that the bar code is read.
Check the screen to verify reagent data is added to the selected position. If
the bar code does not scan, manually type in the information.
6
Close the reagent carousel door.
7
If multiple positions are selected to load or unload, repeat Step 4-6.
8
Before processing samples, calibrate and run QC if needed.
Step
Action
, continued
(2 of 2)
A015924P.EPS
1
3
2