Common Error Messages and Corrective Actions
T
roub
leshooting Calibr
a
tion and Result Errors
UniCe
l DxC Systems Instr
u
ctions F
o
r Use A13914
13-36
October 2005
0x02620019
CC Sample
Delivery
Subsystem Motion
Error, Pausing
CAU: CC Sample
Delivery Subsystem
Motion Error, Pausing.
Message ID:
0x02620019
Subsystem: ccScheduler
Description: This occurs after two
motion errors on the Sample
Delivery Subsystem. Level Sense
errors will not trigger this error. The
system goes into "Pausing" and
then "Stopped" state.
• Raise canopy and check for obvious.
obstruction of the sample probe or
syringe.
• Press the STOP button and HOME the
system. While system is homing,
observe motion of the CC sample
syringe and CC Sample probe. The
probe and syringe should complete
their operations in a smooth manner.
• If the error continues, check the event
log to determine if the syringe or probe
is responsible for the error.
• If the CC sample probe is responsible,
perform the appropriate vertical or
horizontal alignment.
• If the Syringe is responsible, check for
proper seating of syringe assembly.
• Contact Beckman Coulter Technical
Support Center for additional
information.
Table 13.10
Error Messages and Corrective Actions