Chemistry Bypass
Introduction
UniCel DxC Systems Instructions For Use A13914
Reagent Load/Calibration
October 2005
Page 5-27
5
Chemistry Bypass
Introduction
In the event of a calibration failure for a chemistry which has been previously sample
programmed, Chemistry Bypass permits the continuation of analysis of all other
programmed tests without having to deprogram a failed chemistry. The chemistry can
be bypassed while the system status is Standby, Disabled or Running.
• The system will not aspirate sample or dispense reagents for bypassed chemistries.
They will not be marked as Pending.
• Bypassed chemistries are logged on the Post-Run Summary Report.
• Bypassed chemistries must be calibrated to process samples.
Bypassing a Chemistry
The following table explains how to bypass a chemistry.
Removing a Bypassed Status
Recalibrate the chemistry to remove a bypassed status.
Step
Action
1
Select the Rgts/Cal icon from the menu bar.
2
Select the Chem position to bypass.
3
Select Options [F6].
4
From the Cal Options dialog box, select <8> Bypass chemistry.
OR
Type 8 in the Option Number field and press [Enter].