![Beckman Coulter UniCel DxC Synchron 600 Скачать руководство пользователя страница 247](http://html.mh-extra.com/html/beckman-coulter/unicel-dxc-synchron-600/unicel-dxc-synchron-600_instructions-for-use-manual_2719768247.webp)
Overview
Introduction
UniCel DxC Systems Instructions For Use A13914
Maintenance
October 2005
Page 10-1
10
CHAPTER 10 Maintenance
Maintenance Overview
Overview
Introduction
Maintenance on the UniCel DxC System should be performed by trained personnel to
insure continued, optimal performance and to reduce unnecessary service calls.
Minimum training requires that the operator read the Maintenance procedures.
The Scheduled Maintenance Log feature allows the system to keep track of what
maintenance is needed, when it is needed, and who performed the procedure. It also
provides the capabilities of alerting the operator to the maintenance that is due, and to
the maintenance procedures that have been performed and reviewed.
Basic Laboratory Practice
The maintenance frequencies established in this manual are based on processing
approximately 400 samples/day.
The number of samples processed by the system, combined with the nature of the
samples, may require the stated maintenance frequencies to be adjusted in some
laboratories.
A start-up prime of the various subsystems may be necessary if the system has been
idle for more than approximately 8 hours. Refer to
for priming
procedures. Reviewing reagent status prior to running the instrument is standard
laboratory practice.
As with any laboratory equipment, the exterior of the DxC system should be kept
clean. Any spills should be cleaned up promptly. Spills that may be caused by leaking
reagent lines or fittings should be investigated. Refer to
in this chapter for cleaning procedures.
Certain procedures in this manual require the use of a grounded wrist strap while the
procedure is being performed. The strap portion should be placed around the wrist.
The wire portion can be connected to the system in one of the following ways:
• Alligator Clipped: Wrist straps with an "alligator" clip should be clamped onto the
metal frame of the system. To access, open either the left or center door. If desired,
the doors may be closed after the clip is attached to the metal frame.
• Banana Plug Insert: Wrist straps with a "banana" plug must be inserted into the jack
found next to the Stop button on the front of the system.