![Beckman Coulter UniCel DxC Synchron 600 Скачать руководство пользователя страница 282](http://html.mh-extra.com/html/beckman-coulter/unicel-dxc-synchron-600/unicel-dxc-synchron-600_instructions-for-use-manual_2719768282.webp)
Monthly Maintenance
Replace CTS Blade (Sarstedt S-Monovette Type CTS Option only)
Maintenance
UniCel DxC Systems Instructions For Use A13914
Page 10-36
October 2005
8
Carefully move the points of the blade down through the wash tower and
through the alignment slot at the bottom:
• Push down on the blade holder screws and align them over their holes.
• Slowly lower the blade until the blade holder screws are in their holes.
Do NOT hit the points of the blade against the opening as you lower the
blade.
• Install and tighten the two captive blade holder screws. Tighten the
captive blade screw.
9
Install the Cap Piercer cover and tighten the two captive Cap Piercer cover
screws.
10
Select <OK>.
• The system sets the blade count to zero.
• The Cap Piercer carriage moves up to home position.
11
Simulate the load of CTS Auto-Gloss Lubricant by clearing and resetting
the values on the Auto-Gloss screen at <Reagent Load>.
From the MC Reagent Load screen, select Auto Gloss, then [F1] Clear.
Type the following:
• Lot: XXXXX (Refer to Note below)
• SN: NA
• Date: Current month, 1 year greater than the current year
Note: Use 00001 as the first lot number. For each additional load, use one
number greater than the number displayed in the lot number field.
Step
Action
, continued
(4 of 4)