![Beckman Coulter UniCel DxC Synchron 600 Скачать руководство пользователя страница 29](http://html.mh-extra.com/html/beckman-coulter/unicel-dxc-synchron-600/unicel-dxc-synchron-600_instructions-for-use-manual_2719768029.webp)
Operational Conditions
Power Requirements
UniCel DxC Systems Instructions For Use A13914
System Description
October 2005
Page 2-3
2
Power Requirements
Notes on the System Power
The system can operate from any standard 3-wire electrical outlet and is wired as
shipped from the factory to operate on 220V AC, 50/60 Hz.
Environmental Conditions
Table 2.3
Power Requirements
Item
Requirement
Operating range
180–264 V AC, RMS
220 V AC nominal, approximately
15 A at low line, exclusive of power on surge
Frequency
47–63 Hz (50/60 Hz nominal)
BTU generated
DxC 600: 7,200 BTU/hour at nominal (estimated)
DxC 800: 7,500 BTU/hour at nominal (estimated)
DxC PC: 979 BTU/hour
Power connector
20 A current rating. Nema L6-20R twistlock
Uninterrupted power supply Refer to UPS manufacturer’s literature for
specifications
NOTICE
Line Voltage from the electrical outlet should be free of spikes, fluctuations,
and dropouts for protection of the electronic circuitry.
CAUTION
Only operate the system from a 3-wire power source. DO NOT use a 2-prong
adapter or a 2-wire AC power source.
Table 2.4
Temperature, Humidity and Elevation
Item
Specification
Environment
Indoor use only
Ambient temperature
+18°C to +32°C
Warm-up time
30 minutes (time to reach operating temperature)
Relative humidity
20–85% relative, non-condensing
Elevation
Up to 1,280 m (4,200 ft.)