How to Use Reserved Racks
Processing Bar Coded Samples
UniCel DxC Systems Instructions For Use A13914
Preparing Samples for Analysis
October 2005
Page 3-11
3
Processing Bar Coded Samples
Generally, previous sample programming will not need to be cleared before placing a
bar coded sample on the instrument. If the lab reuses Sample IDs, previous
programming must be cleared.
Step
Action
1
Load tubes in a rack with the bar code labels visible through the slot on the
same side of the rack as the rack bar code label.
2
For closed tube sampling systems, check the top of the validated closed
tube for visible blood. If blood is present, clean the rubber stopper with a
cotton tipped applicator stick.
For non-CTS Systems or tubes not validated for closed tube sampling,
remove the cap.
3
For a Routine Sample or if system is in Standby: Priority samples
should be placed on the autoloader first.
• Place the rack in the autoloader with the rack bar code label to the right
and
• Press RUN.
For a STAT Sample: If the system is running and there are other racks on
the autoloader,
• Press PRIORITY. The rack pusher moves back one space so the STAT
rack can be placed in front of the other racks.
• Press RUN.