Common Error Messages and Corrective Actions
T
roub
leshooting Calibr
a
tion and Result Errors
UniCe
l DxC Systems Instr
u
ctions F
o
r Use A13914
13-38
October 2005
0x0262001D
2
CC Reagent Probe
A Motion Error
CAU: CC Reagent
Probe A Motion Error
Message ID:
0x0262001D
Subsystem: RASched
Action Name: action
name
Description: This error is caused
by a Vertical or Rotary motion
error. This will cause the instrument
to go into a "Stopped" state.
• Raise canopy and check for obvious
obstruction of the CC reagent Probe A.
• Press the STOP button and HOME the
system. While system is homing,
observe motion of the CC reagent
probe A. The probe should complete
its operations in a smooth manner.
• If the error continues, check the event
log to determine the specific subsystem
error.
• If event log indicates a vertical error,
make sure the reagent cartridges on the
carousel are pushed in and seated
properly.
• Perform the appropriate vertical or
horizontal CC reagent Probe A
alignment.
• Contact Beckman Coulter Technical
Support Center for additional
information.