Six-Month Maintenance
Replace Glucose Sensor
UniCel DxC Systems Instructions For Use A13914
Maintenance
October 2005
Page 10-63
10
Draining the Reaction Cup for Sensor Maintenance or Replacement
5
Hold the sensor in a vertical position with the membrane tip pointed
upward. Tap on the side of the sensor body to remove any bubbles from
membrane tip.
6
Remove the ISE module and MC module covers.
7
Push the glucose sensor connector into the empty connector port (1) of the
BUNm/UREAm module until the latch clicks.
Figure 10.40
8
Set the glucose sensor down into an empty area of the cup modules. Allow
the free standing sensor to remain connected but not installed for 24 hours,
being careful that the sensor cable is not pinched or crimped. Reinstall the
MC module and ISE module covers.
9
Proceed to Draining Reaction Cup for Sensor Maintenance or Replacement
below to replace the sensor.
NOTICE
System will not autoprime in
Maintenance
mode. After completing the
maintenance procedure, or if a procedure is aborted, exit the Maintenance
menu to resume autoprime and return system to
Standby
.
Step
Action
, continued
(2 of 2)