![Beckman Coulter UniCel DxC Synchron 600 Скачать руководство пользователя страница 280](http://html.mh-extra.com/html/beckman-coulter/unicel-dxc-synchron-600/unicel-dxc-synchron-600_instructions-for-use-manual_2719768280.webp)
Monthly Maintenance
Replace CTS Blade (Sarstedt S-Monovette Type CTS Option only)
Maintenance
UniCel DxC Systems Instructions For Use A13914
Page 10-34
October 2005
3
CAUTION
The points of the Blade are very sharp and extend below the Wash
Tower. Stay away from the bottom of the Wash Tower.
CAUTION
DO NOT select <OK> when the cover is removed. This would move the
Cap Piercer Carriage up and cause a pinching hazard.
• Fully loosen the two captive Cap Piercer cover screws (1) at the base of
the Cap Piercer cover.
• Lift and remove the cover.
Figure 10.18
Step
Action
, continued
(2 of 4)
A014462P.EPS
1