Weekly Maintenance
Clean Total Protein and Albumin Cup Modules (DxC 800 Only)
UniCel DxC Systems Instructions For Use A13914
Maintenance
October 2005
Page 10-11
10
Cleaning Preparation
Prepare the following solutions before beginning the cleaning procedure:
Cleaning Procedure
1N HCL solution
Combine one part 6N HCL and five parts of deionized
water for cleaning Total Protein cup.
OR
10% Cleaning Solution
Combine one part Wash Concentrate II with nine parts
of deionized water for cleaning Albumin cup.
Step
Action
1
Select:
• the Utils icon from the menu bar,
• <2> Maintenance,
• <4> Cup Maintenance,
• <Total Protein> and <Albumin> buttons.
2
Select <Rinse>.
The cups are drained and rinsed 5 times with water. A Result window
appears. DO NOT close this window at this time.
3
When rinsing is complete, use a transfer pipette to manually add:
• One mL of 1N HCL solution to the Total Protein cup.
• One mL of 10% Cleaning solution to the Albumin cup.
4
Let the cleaning solution sit in the cups for 10 minutes. The stir bars will
continue to stir during this time.
5
At the end of the 10 minute waiting period, select <Close> to exit the
Result window.
6
Select <Rinse>.
The cups are drained and rinsed 5 times with water.
7
When rinsing is complete select <Close> to exit the Result window then
select <Drain>.
8
When draining is complete select <OK> to continue. When activity has
completed select <Close> to exit the Result window, and select <Cancel>
to return to the Maintenance Menu.
(1 of 2)