Sample Programming and Processing
Reserved Racks for HbA1c, or IBCT
UniCel DxC Systems Instructions For Use A13914
Preparing Samples for Analysis
October 2005
Page 3-7
3
Reserved Racks for HbA1c, or IBCT
There are two separate fields for specific tests. One for HbA1c and one for IBCT. If a
rack number is entered into one of these fields, any samples run in that rack will only
function for that specific test. For example, Rack 66 is entered into the Reserved
Racks for HbA1c field. Any sample placed in that rack can only have an HbA1c run
on it. The reason for this is that each test in this group requires some sort of sample
preparation. If these tests are part of a group of other tests that do not require
preparation, they cannot be run at the same time as the others. Using these features
allows the prepared sample to be run in one of the reserved racks and allows the result
of the prepared sample to be merged with the original sample report when all tests are
complete.
The default for each of these features is:
• CTS = OFF,
• Reserved racks = none reserved,
• Obstruction detection = On,
• Reserved racks for HbA1c or IBCT = none reserved.
These features can only be modified when the system is in the Standby or Stopped
state.
NOTICE
Even if the Auto Serum Index feature is enabled, no serum indices will be run
for the pre-treated HbA1c and pre-treated IBCT samples run in these reserved
racks.