Common Error Messages and Corrective Actions
T
roub
leshooting Calibr
a
tion and Result Errors
UniCe
l DxC Systems Instr
u
ctions F
o
r Use A13914
13-52
October 2005
0x04B30010
4
MC Reagent Level
High in Reaction
Cup
Chem Name: chem
id
CAU: Reagent too full
Message ID:
0x04B30010
Cup: chem id
Description: MC sample probe
detected reagent in the cup module
above the expected level.
• Prime cup with reagent and make sure
it drains.
• Failure to drain is either caused by low
vacuum or a valve failure.
• Contact Beckman Coulter Technical
Support Center for additional
information.
0x04B30011
4
MC Reagent Level
Low
Chem Name:
CAU: Reagent too low
Message ID:
0x04B30011
Subsystem:
Chem Name:
Description: This error occurs
when the MC sample probe does
not sense reagent at the proper level
or at all during the sample dispense
into the cup module.
• Check Reagent bottle volume.
• Check Reagent tubing to straw
connection is tight.
• Make sure the straw is properly seated
in the cap and reaches to the bottom of
the reagent bottle.
• Enable module and perform reagent/
water primes to make sure that the cup
is filling and draining properly.
• Remove cover and check for leaks or
presence of fluid at the base of the cup
assembly and/or at the electrode/
detector port.
• Contact Beckman Coulter Technical
Support Center for additional
information.
Table 13.10
Error Messages and Corrective Actions
, continued
ID
Class #
Pop-up
Event Log
Description
Corrective Action