Sample Programming and Processing
Sample Preparation by Container Type
Preparing Samples for Analysis
UniCel DxC Systems Instructions For Use A13914
Page 3-4
October 2005
Sample Preparation by Container Type
shows how to prepare different sample containers.
Table 3.2
Preparation of Sample Containers
If running a sample from a...
Then...
Primary Tube
• Use the sample template in the Appendix to determine
adequate sample volume.
• Remove the stopper.
• For CTS systems, remove the stopper, if not a validated
closed tube.
Secondary Tube
• Remove the cap.
• Determine sufficient volume.
• Check for fibrin or other materials resulting from
storage.
Beckman Coulter
Synchron
Microtube™
• Pipette the sample into a Synchron Microtube™.
• Verify there are no bubbles at the bottom of the tube.
• Place into a 13 × 100 "reserved" rack.
Autoanalyzer
Cup
• Place the cup into a rack.
OR
• Place in 15 × 85 tube (with sample bar code on tube).
• Verify there are no bubbles in sample.
• If cup is placed in a tube with a Bar Coded label a
"reserved rack" must be used.
BD Microtainer
®
• Place Microtainer in adapter P/N 472987.
(refer to figure to the right)
• Verify there are no bubbles in sample.
• If cup is placed in a tube with a Bar Coded
label a "reserved rack" must be used.
0.5 mL Cup Insert
(P/N 476399)
(reusable)
• Place the metal Cup Insert into a 16 × 100 mm rack
(P/N 471921 – blue or 474463 – purple).
• Insert a 0.5 mL Autoanalyzer cup into the Cup Insert.
• Run in the "reserved rack" mode.
E014498L.EPS
0.5 mL
2.0 mL
A011538L.EPS