Four-Month Maintenance
Cleaning MC Reagent Lines
UniCel DxC Systems Instructions For Use A13914
Maintenance
October 2005
Page 10-53
10
15
CAUTION
Remove the Glucose
a
sensor before removing the stir bar. To prevent
damage to the glucose sensor membrane tip, do not insert the stir bar
removal tool, applicator stick, or any other object into the glucose
reaction cup unless the sensor has been removed.
CAUTION
Loosen the BUNm/UREAm electrode retainer nut, approximately one-
half turn, and withdraw electrode slightly before removing the stir
bar. This prevents possible damage to the delicate electrode tip.
Using the stir bar removal tool (1), remove the stir bars from all cups (not
necessary for Glucose if the stir bar was previously cleaned during
maintenance). Clean the cups with cotton swabs moistened with deionized
water. Be sure to moisten the swabs before inserting them into the cups.
Note: If the removed swab from the TPm cup is black, prime TPm again
with the 1N HCL solution. Use additional swabs moistened with 1N HCL
solution followed by swabs moistened with deionized water to repeat the
cup cleaning.
Figure 10.31
16
Clean stir bars with a cotton swab or lintless tissue and deionized water.
Use 1N HCL to remove buildup from CREm and TPm stir bars. Always
rinse the cleaning solution from the stir bars using deionized water.
Step
Action
, continued
(3 of 4)