User-Defined Reagent Setup
Defining a User-Defined Reagent
User Defined Reagents
UniCel DxC Systems Instructions For Use A13914
Page 9-4
October 2005
User-Defined Reagent Setup
Defining a User-Defined Reagent
Editing a User-Defined Reagent
User-defined reagent parameters may be edited if:
• The reagent is not on board.
OR
• The reagent is loaded on board and the system is in Standby or Stopped. To modify
the number of calibrations used, the calibrator assignments must be removed.
Step
Action
1
Select the Setup icon from the menu bar.
2
Select <15> User-Defined Chemistries.
3
Select the Number to be defined.
OR
Type a number in the Option Number field and press [Enter].
4
Select Define [F1].
5
Type information for the user-defined reagent as defined on the following
screens:
• Chemistry Parameters.
• Processing Parameters.
• Error Detection Limits.
6
When complete, select an icon from the menu bar for the next desired
action.
If an area is incomplete or incorrect, a flag will appear indicating the error.
Perform the appropriate action to continue. Refer to
in
this chapter for additional information.
7
Select Restore [F1] to restore the default parameters.
OR
Select Cancel [F9] to cancel and return to previous screen.
OR
Select Done [F10] to save and exit.