Common Error Messages and Corrective Actions
T
roub
leshooting Calibr
a
tion and Result Errors
UniCe
l DxC Systems Instr
u
ctions F
o
r Use A13914
13-32
October 2005
0x01FA000B
4
Wash Concentrate
Reservoir Is Not
Filling At All.
CAU: Wash
Concentrate Reservoir Is
Not Filling At All
Message ID:
0x01FA000B
Description: Wash Concentrate
Reservoir Is Not Filling At All.
• Load a new bottle of Wash
Concentrate.
0x01FA0014
4
Both The Waste B
Exit Sump And The
Collection Bottle
Are Full. Switch
#29,15,9.
CAU: Both The Waste
B Sump And The
Collection Bottle Are
Full.
Message ID:
0x01FA0014
Description: Both The Waste B
Sump And The Collection Bottle
Are Full. Only applies to systems
configured with the Split Waste
Option.
• Empty the external waste B collection
bottle and Resume Waste B.
0x01FA0015
4
External Waste B
Collection Bottle Is
Full. Switch #29.
Waste B Paused.
Replace Waste B
Container and
select Resume in
Instrument
Commands when
ready.
CAU: External Waste B
Bottle Is Full.
Message ID:
0x01FA0015
Description: The Waste B
Container is Full. Only applies to
systems configured with the Split
Waste Option.
• Empty the external waste B collection
bottle and Resume Waste B.