![Beckman Coulter UniCel DxC Synchron 600 Скачать руководство пользователя страница 371](http://html.mh-extra.com/html/beckman-coulter/unicel-dxc-synchron-600/unicel-dxc-synchron-600_instructions-for-use-manual_2719768371.webp)
CTS Wick Replacement Procedure
Remove the Blade Assembly and Wick Clip Assembly
UniCel DxC Systems Instructions For Use A13914
Maintenance
October 2005
Page 10-125
10
4
Remove the old Wick Clip Assembly (1) from Wash Tower (2) by lifting
up on both sides of Wick Clip as shown in
. Make sure that the
old Wick is retained within the Wick Clip. If the wick is NOT in the Wick
Clip, the Wick should be found within the Wash Tower area. In this case,
use tweezers to remove the Wick. Put the old Wick and Wick Clip into a
biohazard waste container.
CAUTION
Observe all laboratory practices or procedures that pertain to
handling of infectious and pathogenic materials.
Figure 10.84
Step
Action
(3 of 3)
A007559P.EPS
2
1
1
3