![Beckman Coulter UniCel DxC Synchron 600 Скачать руководство пользователя страница 482](http://html.mh-extra.com/html/beckman-coulter/unicel-dxc-synchron-600/unicel-dxc-synchron-600_instructions-for-use-manual_2719768482.webp)
Common Error Messages and Corrective Actions
T
roub
leshooting Calibr
a
tion and Result Errors
UniCe
l DxC Systems Instr
u
ctions F
o
r Use A13914
13-62
October 2005
0x139D0044
7
Date May Be Out
Of Order Due to
Date/Time Change
SS: date/time was set
backwards from
previous value to new
value.
Description: Information to user
that changing system date or time
(specifically backwards) will cause
certain data to be out of order.
• No operator action is necessary. Be
aware events will be out of order.
0x139F0036
9
One or more
chemistries will
timeout soon
STM: One or more
chemistries will timeout
soon
Description: At least one chemistry
will require recalibration within 1
hour.
• Check Reagent Load/Calibration
screen for calibration information.
• Begin to prepare to recalibrate required
chemistries.
0x139F0037
9
Within Lot Cal
Factor Timeout in
hours
Hours:
Chemistry: chem,
Lot No: lot no,
Serial No: serial
no,
Pos: position
STM: Within Lot Cal
Factor Timeout in hours
Hours:
(Chemistry: chem,
Lot No: lot no,
Serial No: serial no,
Pos: position)
Description: The within lot
calibration factor for the specified
chemistry will timeout soon.
• No operator action is necessary. The
next time a fresh cartridge of the
specified chemistry is loaded, it will
require calibration.
0x139F0038
9
Within Lot Cal
Factor Timeout in
hours
Hours:
Chemistry: chem,
Lot No: lot no,
Serial No: serial no
STM: Within Lot Cal
Factor Timeout in hours
Hours:
(Chemistry: chem,
Lot No: lot no,
Serial No: serial no)
Description: The within lot
calibration factor for the specified
chemistry will timeout soon.
• No operator action is necessary. The
next time a fresh cartridge of the
specified chemistry is loaded, it will
require calibration.
Table 13.10
Error Messages and Corrective Actions
, continued
ID
Class #
Pop-up
Event Log
Description
Corrective Action