![Beckman Coulter UniCel DxC Synchron 600 Скачать руководство пользователя страница 450](http://html.mh-extra.com/html/beckman-coulter/unicel-dxc-synchron-600/unicel-dxc-synchron-600_instructions-for-use-manual_2719768450.webp)
Common Error Messages and Corrective Actions
T
roub
leshooting Calibr
a
tion and Result Errors
UniCe
l DxC Systems Instr
u
ctions F
o
r Use A13914
13-30
October 2005
0x01FA0003
4
DI water reservoir
is not filling at all.
CAU: DI water
reservoir is not filling at
all.
Description: Whenever the DI
water reservoir does not fill within
its given amount of time, it will
time out and flag that it has not
filled.
• Make sure DI water is turned on at the
source.
• Make sure the Manual water inlet
valve is turned on.
• Make sure the instrument has not
rolled on top of the water inlet lines.
• HOME and restart the system.
• If the error continues, contact Beckman
Coulter Technical Support Center for
additional information.
0x01FA0004
4
Flood In Waste
Collection Sump.
Switch #11.
CAU: Flood In Waste
Sump.
Message ID:
0x01FA0004
Description: This error occurs
when the waste collection sump
float switch has indicated that the
canister is full, which is considered
a flood for this canister.
• Check the waste collection sump
canister for liquid. Empty if necessary.
• Press the STOP button and HOME the
system.
• Contact Beckman Coulter Technical
Support Center for additional
information.
0x01FA0006
4
Flood In Primary
Vacuum
Accumulator.
Switch #22.
CAU: Flood In Primary
Vacuum Accumulator.
Message ID:
0x01FA0006
Description: This error will occur
when the Vacuum Accumulator
Float Switch indicates that the
canister is full/flooded. This is a
safety mechanism that prevents a
flood from getting into the pumps.
The Vacuum accumulator should
never have any fluid in it and
should be emptied whenever fluid is
found during Maintenance.
• Check the vacuum accumulator
canister for liquid. Empty if necessary.
• Press the STOP button and HOME the
system.
• If the error continues, contact Beckman
Coulter Technical Support Center for
additional information.
Table 13.10
Error Messages and Corrective Actions
, continued
ID
Class #
Pop-up
Event Log
Description