QC Summary
Introduction
Quality Control
UniCel DxC Systems Instructions For Use A13914
Page 6-16
October 2005
QC Summary
Introduction
• The QC Summary report contains the mean, SD, CV and number of results (N) for
any control run within a specified date interval. The printed report contains the
cumulative mean, SD, CV and (N)umber of accumulated results.
• The QC Summary is also available in the Inter-Laboratory format. The constituent
code must be defined in order for a chemistry to be included in the Inter-Laboratory
QC Summary. To enter the Inter-Laboratory reference number and contact person to
be included in the Inter-Laboratory QC Summary, access Report Setup through
System Setup.
Accessing the QC Summary
Follow the steps below to either access or print the QC summary.
Step
Action
1
Select the QC icon from the menu bar.
2
A list of currently defined controls appears. Select the control for the QC
Summary, or type the item number and press [Enter].
3
Select Summary [F6] from the QC screen.
4
Type a Start date and an End date using the dd/mm/yy format. The default
Start/End date is the current date. Press [Enter] to select the default, or
after entering the Start and End dates. Press the [Tab] key to move
between fields.
Select <OK> to view the QC Summary.
OR
Select <Cancel> to return to QC screen.
If no data is available, the following message appears: "No Data
Available."
5
To print a hard copy of the summary, select Print [F9]. To print the Inter-
Laboratory format of the QC Summary, select Inter Lab [F8].
6
To exit the QC Log Screen, select Done [F10].