Six-Month Maintenance
Replace Calcium and Potassium Electrode Tip
Maintenance
UniCel DxC Systems Instructions For Use A13914
Page 10-60
October 2005
3
Locate Potassium electrode (1). Remove electrode by turning the electrode
retaining nut counterclockwise.
AND/OR
Locate Calcium electrode (2). Remove electrode by turning the electrode
retaining nut counterclockwise.
Figure 10.38
4
Remove the retainer nut from the old electrode. Remove the quad-ring (1)
from the tip of the old electrode. Inspect the electrode port if the quad-ring
is not on the electrode. Discard the quad-ring.
Figure 10.39
5
Unscrew the old tip from the electrode assembly. Discard the O-ring and
the old electrode tip.
Step
Action
, continued
(2 of 4)
U