As-Needed/As-Required Maintenance
EIC Port Cleaning
UniCel DxC Systems Instructions For Use A13914
Maintenance
October 2005
Page 10-91
10
18
Reinstall the solenoid valve to the back side of the EIC. Orient the valve to
align the guide pins (1) on the valve with the appropriate holes (2) on the
EIC. Press the EIC and valve firmly together for a flush fit. Tighten the two
screws on the valve, turning the screws until resistance is felt. At this point,
tighten each screw 1/8 of a turn further. DO NOT overtighten the screws.
Figure 10.68
19
Reattach lines #18, #24, #23, #26, and #15 to the EIC ports. (Refer to
20
Remove clamps from lines #18, #24, #23, and #26.
21
Reattach the EIC top sections to the drip tray. Align the guide pin on the
bottom of the EIC to the hole on the drip tray and press the EIC down onto
drip tray. Tighten the Phillips screw on the top of the EIC.
22
Reconnect the solenoid valves at the valve connector.
23
Select <OK> to close the pop-up Maintenance Confirmation window.
NOTICE
Reagents will immediately prime through the flow cell.
24
Select <Close> in the pop-up Result window to exit the ISE Service
screen.
25
Observe EIC, lines and connections to verify that reagents are flowing
properly with no leaks.
Step
Action
, continued
(6 of 7)