Monthly Maintenance
Replace Alkaline Buffer
Maintenance
UniCel DxC Systems Instructions For Use A13914
Page 10-26
October 2005
3
Press the latch, which is located on the connector (1), to disengage the
electrode pin lead from the connector panel and remove the electrode
connector from the socket.
Figure 10.12
4
Unscrew the BUNm/UREAm electrode retainer nut (1) and withdraw the
electrode-retainer assembly from the chemistry reaction cup.
Figure 10.13
Step
Action
, continued
(2 of 5)