Instrument Function
R&S
®
SMJ100A
271
Operating Manual 1403.7458.32 ─ 14
"Extern Start/Stop"
Sets an automatically repeated sweep cycle that is started, stopped
and restartet by subsequent external trigger events.
The first external trigger signal starts the sweep (Start).
The next external trigger signal stops the sweep at the current fre-
quency (Stop).
The third external trigger signal starts the sweep at the start fre-
quency (Start).
Refer to the description of the rear panel for information about the
connectors for external trigger signal input (see
Example:
SOUR:SWE:POW:MODE AUTO
TRIG:PSW:SOUR EAUT
SOUR:POW:MODE SWE
(External trigger)
Remote command:
[:SOURce<hw>]:SWEep:POWer:MODE
Reset Sweep - Level Sweep
Resets the sweep. The start level is set and the next sweep starts from there.
Remote command:
[:SOURce<hw>]:SWEep:RESet[:ALL]
Execute Single Sweep - Level Sweep
Triggers the sweep manually. A manual sweep can only be triggered if "Mode Single"
is selected.
Example:
SOUR:SWE:POW:MODE AUTO
TRIG:PSW:SOUR SING
SOUR:POW:MODE SWE
SOUR:SWE:EXEC
Remote command:
[:SOURce<hw>]:SWEep:POWer:EXECute
:TRIGger<hw>:PSWeep[:IMMediate]
:TRIGger<hw>[:SWEep][:IMMediate]
Start Level - Level Sweep
Sets the start level.
Remote command:
RF Signal and Analog Modulations - RF Block
Summary of Contents for SMJ100A
Page 34: ...Preface R S SMJ100A 18 Operating Manual 1403 7458 32 14 Notes on Screenshots...
Page 86: ...Preparing for Use R S SMJ100A 70 Operating Manual 1403 7458 32 14 LXI Configuration...
Page 826: ...Remote Control Commands R S SMJ100A 810 Operating Manual 1403 7458 32 14 UNIT Subsystem...
Page 828: ...Maintenance R S SMJ100A 812 Operating Manual 1403 7458 32 14 Storing and Packing...
Page 844: ...Hardware Interfaces R S SMJ100A 828 Operating Manual 1403 7458 32 14 BERT Connector...