Remote Control Commands
R&S
®
SMJ100A
795
Operating Manual 1403.7458.32 ─ 14
Example:
SYSTem:SERRor?
queries all errors existing in the error queue.
Response: -221, 'Settings conflict', 153,
'Input voltage out of range'
the two returned errors have occurred since the error queue was
last queried.
Usage:
Query only
:SYSTem:COMMunicate:GPIB:LTERminator
<LTerminator>
Sets the terminator recognition for remote control via GPIB bus.
Parameters:
<LTerminator>
STANdard | EOI
EOI
The terminator must be sent together with the line message EOI
(End of Line). This setting is recommended for binary block
transmissions where a character could coincidentally have the
value LF (Line Feed) but is not intended as the terminator. This
setting must be selected for block data with undefined length.
STANdard
An LF (Line Feed) is recognized as the terminator regardless of
whether it is sent with or without EOI.
*RST:
STANdard
Example:
SYSTem:COMMunicate:GPIB:LTERminator EOI
only a character which is sent simultaneously with the line mes-
sage EOI is accepted as the terminator.
:SYSTem:COMMunicate:GPIB[:SELF]:ADDRess
<Address>
Sets the GPIB address.
Parameters:
<Address>
integer
Range:
1 to 30
*RST:
28
Example:
SYSTem:COMMunicate:GPIB:SELF:ADDRess 28
sets GPIB address.
Manual operation:
See
:SYSTem:COMMunicate:NETWork[:COMMon]:WORKgroup
<Workgroup>
Sets the individual workgroup name of the instrument.
Parameters:
<Workgroup>
string
SYSTem Subsystem
Summary of Contents for SMJ100A
Page 34: ...Preface R S SMJ100A 18 Operating Manual 1403 7458 32 14 Notes on Screenshots...
Page 86: ...Preparing for Use R S SMJ100A 70 Operating Manual 1403 7458 32 14 LXI Configuration...
Page 826: ...Remote Control Commands R S SMJ100A 810 Operating Manual 1403 7458 32 14 UNIT Subsystem...
Page 828: ...Maintenance R S SMJ100A 812 Operating Manual 1403 7458 32 14 Storing and Packing...
Page 844: ...Hardware Interfaces R S SMJ100A 828 Operating Manual 1403 7458 32 14 BERT Connector...