Remote Control Commands
R&S
®
SMJ100A
557
Operating Manual 1403.7458.32 ─ 14
Example:
OUTP:PROT:TRIP
Queries the state of the protective circuit for RF output A.
Response: 0
The protective circuit has not tripped.
Response: 1
The protective circuit has tripped.
Usage:
Query only
Manual operation:
See
:OUTPut<hw>[:STATe]
<State>
Activates and deactivates the RF output signal (RF ON / RF OFF).
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
0
Example:
OUTP OFF
deactivates the RF output.
Manual operation:
See
:OUTPut<hw>[:STATe]:PON
<Pon>
Selects the state of the RF output when the instrument is switched on.
Parameters:
<Pon>
OFF | UNCHanged
OFF
deactivates the output when the instrument is switched on (RF
OFF).
UNCHanged
restores the initial state of the RF output before the last turn off.
sets the output status as it was when the instrument was
switched off.
*RST:
UNCHanged
Example:
OUTP:PON OFF
RF output A is deactivated when the instrument is switched on.
Manual operation:
See
:OUTPut:USER<ch>:SOURce
<Source>
The command selects the signal for the specified USER interface.
Some signals which can be applied at the USER interface are permanently assigned
and some are assigned using a dedicated command (e.g.
OUTP:CW:SOURce
for the
CW control signal).
OUTPut Subsystem
Summary of Contents for SMJ100A
Page 34: ...Preface R S SMJ100A 18 Operating Manual 1403 7458 32 14 Notes on Screenshots...
Page 86: ...Preparing for Use R S SMJ100A 70 Operating Manual 1403 7458 32 14 LXI Configuration...
Page 826: ...Remote Control Commands R S SMJ100A 810 Operating Manual 1403 7458 32 14 UNIT Subsystem...
Page 828: ...Maintenance R S SMJ100A 812 Operating Manual 1403 7458 32 14 Storing and Packing...
Page 844: ...Hardware Interfaces R S SMJ100A 828 Operating Manual 1403 7458 32 14 BERT Connector...